IDRaman reader

Integrated Raman Spectrometer

Please contact us online for more information about ordering this product.

View Product Variations

The IDRaman reader from Ocean Optics is a fully integrated Raman spectrometer system featuring Raster Orbital Scanning technology (ROS). Scanning a tightly focused laser beam over the sample surface maximizes sensitivity and maintains high resolution for the most reliable Raman measurements. A convenient sampling system allows you to measure surfaces below the instrument, in cuvettes, or from the side and bottom of vials. This configuration is ideal for process Raman measurements or reading SERS substrates.

  • Laser excitation wavelengths – 532, 638, 785, or 808 nm
  • Raster Orbital Scanning — novel sampling technique improves measurement repeatability and decreases risk to samples
  • Powerful – high-sensitivity Raman measurements
  • Flexible – measure both liquid and solid samples

Engineering Specifications IDRaman reader (638 nm, 785 nm and 808 nm versions) IDRaman reader (785 nm and 808 nm high resolution versions) IDRaman reader (532 nm version)
PHYSICAL
Dimensions: (LWH) mm and inches: 28.2 cm (9 in.) x 14 cm (5.5 in.) x 6.4 cm (2.5 in.) 28.2 cm (9 in.) x 14 cm (5.5 in.) x 6.4 cm (2.5 in.) 30.5 cm (12 in.) x 20.3 cm (8 in.) x 7.6 cm (3 in.)
Weight: 2.7 kg (6 lb.) 2.7 kg (6 lb.) 3.6 kg (8 lb.)
SYSTEM
Raman laser excitation wavelengths: 638 nm, 785 nm and 808 nm 785 nm and 808 nm 532 nm
Laser output power: 100 mW (~70 mW at sample) 100 mW (~70 mW at sample) 50 mW (~45 mW at sample)
Range/resolution: 200-3200 cm-1 200-2000 cm-1 200-3200 cm-1
Resolution: 10 cm-1 4 cm-1 7 cm-1
Detector: 2048-element NIR-enhanced CCD array with thermoelectric cooling to 10 °C
Power: 12 VDC with supplied 110-240V AC power supply
Optics: NA = 0.5, focal length of 11 mm and 2 µm spot size (2 mm with rastering engaged)
Raster Orbital Scanning: sampling function allows detection of inhomogeneous samples without sample degradation
Sampling options: Vial (square or round) through side or bottom; downward-facing free space
Accessories: Stage with 3-axis manual adjustment; USB camera option for recording free space sampling

Measurement Techniques in which this product is used: